Patterns of adaptive servo-ventilation settings in a real-life multicenter study: pay attention to volume!

Résumé

Backgrounds: To explain the excess cardiovascular mortality observed in the SERVE-HF study, it was hypothesized that the high-pressure ASV default settings used lead to inappropriate ventilation, cascading negative consequences (i.e. not only pro-arrythmogenic effects through metabolic/electrolyte abnormalities, but also lower cardiac output). The aims of this study are: i) to describe ASV-settings for long-term ASV-populations in real-life conditions; ii) to describe the associated minute-ventilations (MV) and therapeutic pressures for servo-controlled-flow versus servo-controlled-volume devices (ASV-F Philips®-devices versus ASV-V ResMed®-devices). Methods: The OTRLASV-study is a cross-sectional, 5-centre study including patients who underwent ASV-treatment for at least 1 year. The eight participating clinicians were free to adjust ASV settings, which were compared among i) initial diagnosed sleep-disordered-breathing (SBD) groups (Obstructive-Sleep-Apnea (OSA), Central-Sleep-Apnea (CSA), Treatment-Emergent-Central-Sleep-Apnea (TECSA)), and ii) unsupervised groups (k-means clusters). To generate these clusters, baseline and follow-up variables were used (age, sex, body mass index (BMI), initial diagnosed Obstructive-Apnea-Index, initial diagnosed Central-Apnea-Index, Continuous-Positive-Airway-Pressure used before ASV treatment, presence of cardiopathy, and presence of a reduced left-ventricular-ejection-fraction (LVEF)). ASV-data were collected using the manufacturer’s software for 6 months. Results: One hundred seventy-seven patients (87.57% male) were analysed with a median (IQ25–75) initial Apnea-Hypopnea-Index of 50 (38–62)/h, an ASV-treatment duration of 2.88 (1.76–4.96) years, 61.58% treated with an ASV-V. SDB groups did not differ in ASV settings, MV or therapeutic pressures. In contrast, the five generated k-means clusters did (generally described as follows: (C1) male-TECSA-cardiopathy, (C2) male-mostly-CSA-cardiopathy, (C3) male-mostly-TECSA-no cardiopathy, (C4) female-mostly-elevated BMI-TECSA-cardiopathy, (C5) male-mostly-OSA-low-LVEF). Of note, the male-mostly-OSA-low-LVEF-cluster-5 had significantly lower fixed end-expiratory-airway-pressure (EPAP) settings versus C1 (p = 0.029) and C4 (p = 0.007). Auto-EPAP usage was higher in the male-mostly-TECSA-no cardiopathy-cluster-3 versus C1 (p = 0.006) and C2 (p < 0.001). MV differences between ASV-F (p = 0.002) and ASV-V (p < 0.001) were not homogenously distributed across clusters, suggesting specific cluster and ASV-algorithm interactions. Individual ASV-data suggest that the hyperventilation risk is not related to the cluster nor the ASV-monitoring type. Conclusions: Real-life ASV settings are associated with combinations of baseline and follow-up variables wherein cardiological variables remain clinically meaningful. At the patient level, a hyperventilation risk exists regardless of cluster or ASV-monitoring type, spotlighting a future role of MV-telemonitoring in the interest of patient-safety.
Fichier principal
Vignette du fichier
2020 Jaffuel et al., Patterns of adaptative.pdf (961.95 Ko) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Loading...

Dates et versions

hal-02946896 , version 1 (24-09-2020)

Identifiants

Citer

Dany Jaffuel, Claudio Rabec, Carole Philippe, Jean-Pierre Mallet, Marjolaine Georges, et al.. Patterns of adaptive servo-ventilation settings in a real-life multicenter study: pay attention to volume!. Respiratory Research, 2020, 21, pp.243. ⟨10.1186/s12931-020-01509-7⟩. ⟨hal-02946896⟩
55 Consultations
86 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More