Abstract : It is important to size and optimize the testing chamber in order to ensure a qualified sensor response measurement. Indeed, the amplitude, as well as the response and recovery times are very dependent on the testing chamber. In this study, we show that by optimizing the design and reducing the volume of the testing chamber, the responses of metal oxide microsensors are highly enhanced and faster, and therefore closer to the real behavior of the sensor.
https://hal-amu.archives-ouvertes.fr/hal-02407313
Contributor : Gilles Bouchet <>
Submitted on : Thursday, December 12, 2019 - 2:30:57 PM Last modification on : Wednesday, December 18, 2019 - 1:49:56 AM Long-term archiving on: : Friday, March 13, 2020 - 7:28:00 PM